Xcite Stimulator and cart

Size and weight

Length 17” (84cm)
Width  20” (50cm)
Height 33” (92cm) with Xcite Stimulator 39” (99cm)
40lb (18kg)

Maximum load rating of cart

25kg (55lb)

Power supply

100-240VAC, 50/60Hz, 1.3A

Key materials

Steel, HDPE

Protective type

II

Degree of protection

Type BF Applied Part

Noise emission

No moving parts or significant noise emission

Mode of operation

This device is for continuous use

Conditions:

 

Operating

5°C to 35°C (41°F to 95°F),
15% to 93% Rh, non-condensing
700hPa to 1060hPa
Maximum altitude 6,561ft (2,000m)

Maximum applied parts temperature

50°C (122°F) (touchscreen)
42°C (108
°F) (enclosure top)

storing

Between -20°C (-4°F), without relative humidity and +45°C (+113°F), at relative humidity up to 93% (non-condensing)

Xcite Stimulator

 

Display / interface

10", touch sensitive LCD

Operating system

Windows 10 Professional

Internet access for RTILink.com

Requires access to https://www.rtidatalink.com

LAN connection

Ethernet LAN 10/100Base-T, RJ45 connector

Stimulator

 

Maximum voltage output

200V

Maximum number of channels

12

Current output per channel

0 – 140mA (± 2mA) with 100 to 1500W load

Waveform type

Alternating monophasic, charged balanced

Pulse period

10 – 100msec; default: 30msec

Pulse width

50 – 500ms; default: 250ms

Maximum output charge per pulse frequency

70mC

Graphical representation of typical output

Pulse Train.jpg